EXT-16DX

EXT-Series

Description

Features:

  • Extracellular recordings with up to 16 channels
  • Up to 16 gain stages, 16 high pass and 16 low pass filter corner frequencies; set by rotary switch
  • Customized configuration for all channels (set by factory)
  • Digital display of selected channel, gain, high pass and low pass filter corner frequency
  • Remote operation with digital I/O card or manual operation with switches and push button
  • Brightness of displays adjustable for simultaneous fluorescence measurements
  • Referenced or single ended recording mode (set by switch)
  • External npi power supply or rechargeable batteries for extra-low-noise operation
  • Extracellular signals available at BNC connectors or at SCSI/SubD connector for direct linking to data acquisition card

With R/I-T1DX

  • Electrode resistance test separately for each electrode
  • Current injection (up to ±1 µA, bipolar) for each electrode

Headstages

EXT-MINI-16-DIL

16-channel headstage with current injection

  • 16 single-ended channels
  • One layer for current injection / resistance measurement
  • Shielding layer
  • Designed for use with NeuroNexus probes with A16 package

EXT-MINI-16-STIM

Switchable 16-channel stimulation headstage

  • 16 single-ended recording channels
  • Each channel switchable to high-voltage stimulation
  • 1mm pins for connecting electrodes

EXT-MINI-16-ACC

16 channel headstage for in-vivo recordings

  • 16 recording channels
  • Built-in acceleration sensor (3 axis)
  • Two independent stimulation lines
    for high-voltage stimulation with separate electrodes
  • Omnetics connector for direct connection of
    NeuroNexus probes with CM-16 package

Technical Data

Input:    
>1012 Ω, range ±2 V

Differential Input:
CMR >70 dB at 1 kHz (tested with 0 Ω input resistance)

Input Capacitance:
10 pF

CHANNEL Monitor:
-8V…+7V; +13V in single ended mode

HIGHPASS Filter:
single pole, attenuation: -6 dB / octave, corner frequencies (Hz): DC, 0.1, 0.3, 0.5, 1, 3, 5, 10, 30, 50, 100, 300, 500, 800, 1k, 3k; selected by rotary switch

HIGHPASS Monitor: 
-8V…+7V

LOWPASS Filter: 
single pole, attenuation: -6 dB/octave, corner frequencies (Hz): 20, 50, 100, 200, 300, 500, 700, 1k, 1,3k, 2k, 3k, 5k, 8k, 10k, 13k, 20k; selected by rotary switch

LOWPASS Monitor:
-8V…+7V

GAIN:
10, 20, 50, 100, 200, 500, 1k, 2k, 5k, 10k; selected by rotary switch

GAIN Monitor:
+1V…+10V

Output:
range: ±10 V into 1 kW / ±1 V into 50 W

External power supply:
115/230 Volt AC, 60/50 Hz, fuse 1.6/0.8 A, slow

Dimensions:
19” rackmount cabinet
19” (483 mm), 10” (250 mm), 3.5” (88 mm)

Headstage Size:
without housing, approx. 40 mm x 18 mm, with holding bar

Downloads

Manuals:
EXT-02F Manual
EXT-02B Manual
EXT-T2 Manual
EXT-16DX Manual

Brochure:
EXT Brochure

References:
EXT-Reference List
EXT-02F Reference List
EXT-16DX Reference List