MVCS + MVCC Modular

MVCS Series – Iontophoresis Amplifiers

Description

MVCS Series – Iontophoresis Amplifiers
Features:

  • Time resolution: down to 100 µs, spatial resolution: down to 1 µm
  • Simulation of synaptic events, suitable for receptor density mapping
  • Automated balancing of iontophoretic current (option)
  • Currents from tens of pA up to µA
  • High-voltage, high-speed current source
  • Automated electrode resistance test

Works only with EPMS-H-07!

Ordering

Part no. Description  
MVCS-01M-45 Standard system One channel system without capacity compensation, width = 1.5
MVCS-02M-45 Standard system Tow channel system without capacity compensation, width = 3
MVCC-M-45 Standard system Balance module for up to four channels without capacity compensation, width = 1
MVCS-C-01M-45 Very fast system One channel system with capacity compensation, electrode resistance test and a headstage, width = 1.5
MVCS-C-02M-45 Very fast system Dual channel system with capacity compensation, electrode resistance test and two headstages, width = 3
MVCC-C-M-45 Very fast system Balance module for up to four channels with capacity compensation, electrode resistance test and a headstage, width = 1
EH-01 Accessories Microelectrode holder for sharp microelectrodes, 1.5 mm standard, without ports, BNC

Technical Data

HEADSTAGES:
Size (approx.): Standard 65x25x25 mm, holding bar diameter 8 mm, length 150 mm, mounting plate or dovetail on request,
Electrode connector: BNC connector,
Ground connector: 2.4 mm connector.
Electrode output: floating current source, output impedance >1012 Ω.

MAXIMUM CURRENT:
±450 nA (±45 V) or ±1.5 µA (±150 V) into 100 MΩ load. Other ranges available on request.

DISPLAY:
current XXXX nA, voltage XXX.X V, REL XXXX MΩ,
separate displays for each channel, displayed value is set by a three position toggle switch.

OVER LEDs:
activated 10 % below maximum current.

EJECT:
ten-turn control, switch selected range max. 1 µA / 100 nA.

RETAIN:
ten-turn control, max. 100 nA.

CAPACITY COMPENSATION:
range 0 ­ 30 pF, ten-turn control.

OUTPUT CURRENT POLARITY:
selected by INVERTED / NORMAL toggle switch.

MODE of operation:
set with toggle switches,
EJECT / RETAIN / AUTO switch enables manual or TTL controlled operation,
SET / OPERATE switch connects automatically electrode outputs to an internally grounded load, to make possible well defined presetting (SET position),

TTL INPUT (AUTO mode):
LO = RETAIN, HI = EJECT, isolated, Rin > 5 kΩ.

ANALOG INPUT:
sensitivity 100 nA / V, Rin > 100 kΩ, range ±10 V.

CURRENT MONITOR:
sensitivity 100 nA / V, Rout = 50 Ω, not isolated.

VOLTAGE MONITOR:
Vel / 10, Rout = 50 Ω, not isolated.

ELECTRODE RESISTANCE TEST:
obtained by application of square current pulses, ±10 nA from built-in pulse generator.

POWER REQUIREMENTS:
115 / 230 V AC, 45 – 60 W.