PDES – Pressure Ejection

PDES modules for the EPMS system

Description

PDES modules for the EPMS system are available with digital and analog manometer and one channel. They can be operated using an internal or an external valve. Maximum pressure is 4 bar / 58 psi.
By default the PDES modules come with an analog timer. For digital timing an additional module, TMR-02M, is required.

PDES systems are also available as 19″ stand-alone devices.

PDES - Models

pdesmap

Ordering

Part no. Description
PDES-01T-M internal valve, one channel, analog timer, analog manometer, width = 1.5
PDES-01T-M-LA internal valve, one channel, analog timer, digital manometer, width = 1.5
PDES-01T-ME external valve, one channel, analog timer, analog manometer, width = 1.5
PDES-01T-ME-LA external valve, one channel, analog timer, digital manometer, width = 1.5
PDES-EHM remote microJECT holder with one fast valve (< 1 ms)
PDES-01-BOX remote valve box for one slow valve (ca. 10 ms)
PDES-02-BOX remote valve box for two slow valves (ca. 10 ms)
PDES-EH EH-IPH-1-THP Injection pipette holder
PDES-PRS Remote switch push button for PDES modules and DXH systems
PDES-T-4 Tubes for connecting pipettes, outer diameter 4mm; 2 meters length, width 30cm, flexible tubing for connecting the pipette, per channel
PDES-T-6M Pressure tubing for connecting the pressure source to the controller, outer diameter 6 mm, for one channel

Technical Data

Source gas:
air, nitrogen, helium etc. no aggressive, combustible or explosive gases, filtered by a 5 µm filter element

Pressure range:
input pressure:
max. 5 bar (1 bar = 105 Pa)
pressure control range:
0-4 bar; display range 0-4 bar or 0-2 bar; display range 0-2.5 bar

Input and output couplings:
hose fitting connect type for 4 mm (inner diameter) tubing

Status / EJECT time indicator:
red/green LED
red light:    valve is closed (in EXTERN TIMER, CONT. operation mode)
green light:    valve is open
Off (no light)    valve is closed

Mode selection:
four position toggle switch, OFF, EXTERN TIMER, CONT.

Valve switching time:
10-20 ms (dependent on tubing) / <1 ms with microJECT

Analog timer:
0.001 – 0.1 s, 0.01 – 1 s, 0.1 – 10 s or 1 – 100 s with linear control, range set by rotary switch time base selector (x0.01 s, x0.1 s, x1 s, x10 s)

TTL INPUT:
standard BNC connector, TTL (LO = 0-2.4 V, HI = 2.5-10 V), input impedance 10 kΩ