TEC-03X

TEC family

Overview

TEC-03X – Two-Electrode Voltage Clamp Amplifier for routine recordings from oocytes

Features:

  • The TEC-03X is based on the standard two electrode approach and is an ideal, easy-to-use system for recording from oocytes
  • Accurate and fast two-electrode voltage and current clamp (V/C) amplifier with PI-controller for studying large membrane currents
  • Differential potential registration and high-voltage current source output, automated electrode resistance test mode which can be used even with the electrodes impaled in an oocyte
  • Digital DISPLAYS for current, voltage and electrode resistance
  • Two-pole (optionally four-pole) BESSEL filter for current
  • No virtual ground needed for recording membrane currents
  • OSCILLATION SHUT-OFF unit prevents cells from damage
  • Standard current range is ±150 µA into 1 MOhm. Current headstages with selectable current ranges (x0.1, x1, x2, x5 or x0.1, x0.2, x0.5, x1) are also available
  • Easy operation with all major data acquisition systems, remote selection of MODE of OPERATION (CC, VC), telegraphing (monitoring) outputs for current sensitivity and filter
  • Optional: Built-in interface for data acquisition with CellWorks

Features

Feature TEC-03X TEC-05X TEC-10CX
VC-Error display X X
PI-Controller X X X
Series Resistance Compensation X X X
Electrode Resistance Test X X X
Mode of Operation Switch Rotary switch Rotary switch Push buttons
Bridge Mode X
DHC mode opt.
Lowpass filter: 20 … 20 kHz Two-Pole BESSEL
(opt. Four-Pole)
Two-Pole BESSEL
(opt. Four-Pole)
Four-Pole BESSEL
Direct Current output from Headstage X X X
Oscillation Shut-off X X X
Transient Compensation X
Capacity Compensation X X X
Cap. Comp. of Current Headstage X
Slow Mode: VCcCC
(Sutor et al., 2003)
opt.
Please note:
These are the available standard configurations.
If you have special requirements,
please contact us directly.

Ordering

Part no. Description
TEC-03 TEC 03X Oocyte Voltage/Current Clamp Amplifier
TEC-03-BF TEC-03X-BF Oocyte Amplifier with fourpole BESSEL filter
TEC-03-BF-CW TEC-03X-BF Oocyte Amplifier with fourpole BESSEL filter and built-in interface for CellWorks software
TEC-03-CW TEC-03X with built-in interface for CellWorks software
TEC-EH PPH-0P-BNC Microelectrode Holder 1.5 mm glass, w/o. port, w. BNC connector
TEC-EH-SET Electrode Holder Set for TEC Amplifiers
TEC-MOD Passive Cell Model for TEC Systems
TEC-MODA Cell Model with Active Membrane Resistance for TEC systems
TEC-CON SubClic Cable for TEC Headstage

Technical Data

MODES OF OPERATION:
RPel: Potential Electrode Resistance Test;
CC: Current Clamp Mode;
VC: Voltage Clamp Mode;
RCel: Current Electrode Resistance Test;
BR: Bridge Mode (TEC-05X);
DHC Mode (TEC-05X, option);
HEADSTAGES
Potential headstage:
Operating voltage ±15 V;
Size (approx.): 70x26x26mm, holding bar diameter 8 mm, length 150 mm;
Input resistance: >1013 Ω;
Differential input (for supression of bath potentials): cmr >80 dB.

Current headstage (15 V, TEC-05X)):
Operating voltage: ±15 V;
Size (approx.): same as potential headstage,
Input resistance: >1012 Ω (internally trimmable).

Current headstage (150 V):
Operating voltage: ±150 V (standard);
Size (approx.): 100x50x30 mm, grounded enclosure,
holding bar diameter 8 mm, length 10 cm (isolated from ground, standard system only);
Input resistance: >1012 Ω (internally trimmable);
Power dissipation: 6 W (standard)

FEATURES:
Oscillation Shut-off
Electrode Resistance Test
Proportional Integrator
Series Resistance Compensation
Transient Compensation (TEC-10CX)

OUTPUTS:
CURRENT OUTPUTS:
Uncompensated output signal: sensitivity 0.1 V/µA or 0.1 nA/V (TEC-05X), voltage range ±15 V;

Compensated/filtered output: sensitivity: 0.1 V…10 V/µA (V/nA for TEC-05X)
POTENTIAL OUTPUTS:
Potential electrode: two outputs, sensitivity x10 mV only (TEC-03X) or x10 mV and x40 mV (TEC-05X, TEC-10CX), voltage range ±15V;
Current electrode: sensitivity x10 mV, voltage range ±15 V;

INPUTS:  
CURRENT CLAMP (standard headstage):
Inputs: 1 µA/V, 0.1 µA/V with ON/OFF switches (TEC-10CX), 1 µA/V (TEC-03X), 1 nA/V, 0.1 nA/V with ON/OFF switches (TEC-05X);
input resistance >100 kΩ;
VOLTAGE CLAMP:
Inputs: :10 mV and :40 mV (TEC-05X, TEC-10CX) with ON/OFF switches, input resistance >100 kΩ;

CONTROLS:

MODE selection: rotary switches (TEC-03X, TEC-05X), or pushbuttons (TEC-10CX) , LED indicators;

remote selection by TTL inputs, software assessable (optional).
Current Electrode Parameter Controls:
Offset compensation: ten-turn control, ±1200 mV;
Capacity compensation (optional, TEC-05X): range 0 – 30 pF, ten-turn potentiometer.
Potential Electrode Parameter Controls:
Offset compensation: ±200mV, ten-turn control;
Capacity compensation: range 0 – 30 pF, ten-turn control.
Oscillation Shut off: disabled / off / reset switch, threshold set with linear control (0- 1200 mV).
Current Clamp: HOLD: X.XX µA (TEC-03X, TEC-10CX), X.XX nA, ten-turn digital control with -/0/+ switch, maximum 10 µA (TEC-03X, TEC-10CX) or 10 nA (TEC-05X);
BRIDGE balance (optional, TEC-05X systems): max. 10 MΩ or max. 100 MΩ (switch selected) with ten-turn digital control.
INTEGRATOR TIME CONSTANT: 200 µs – 2 ms, ten-turn control (TEC-03X) or control with ON/OFF switch (TEC-05X, TEC-10CX) ;

Voltage Clamp: HOLD: XXX mV, ten-turn digital control with +/0/- switch, maximum 1000 mV;

RISE TIME LIMIT: 0 – 0.2 ms;
GAIN: 10 µA/V – 10000 µA/V, ten-turn linear control;
INTEGRATOR TIME CONSTANT: 200 µs – 2 ms, ten-turn control (TEC-03X) or control with ON/OFF switch (TEC-05X, TEC-10CX) ;

OUTPUT CURRENT LIMIT5 (TEC-10CX): 0 – 100%, linear control;

OUTPUT FILTERS:
Current output: lowpass  filters with 16 corner frequencies (20 Hz – 20 kHz)

AUDIO MONITOR:
Pitch correlated with potential signal.

DISPLAYS:
potential display: XXX mV.
Electrode resistance display XXX MΩ.
current display: X.XX µA
VC error (TEC-10X)

POWER REQUIREMENTS:
115 V/230 V AC, 60 W (1.25 A/0.63 A fuse, SLOW).

DIMENSIONS:

19″ rackmount cabinet, 19″ (483 mm) wide, 14″ (355 mm) deep, 5.25″(132.5 mm) high
weight: approx. 8 kg.

Downloads

Manuals:
TEC-03X Manual
TEC-05X Manual
TEC-10CX Manual

TEC Cell Model Manual
TEC-RS Cell Model Manual

Brochure:
TEC Brochure

HOWTO…

References:
TEC Reference List
TEC-03X Reference List
TEC-05X Reference List
TEC-10CX Reference List