Pneumatic Ejector and Extracellular Amplifier

Pext-01

Add to quote list
This product is already in quote request list
Browse the list
Category: Drug Application via Pressure Ejection (PDES systems) Tags: focal micro perfusion, perfusion outlets, perfusion system, peristaltic pump, superfusion, syringe pumps
  • Overview
  • Details
  • Specifications - PDES
  • Specifications - EXT
  • Downloads
  • Ordering

Overview

The PEXT-01 is designed for pressure ejection of drugs in physiological and pharmacological studies while recording extracellular signals.

It combines the popular PDES pressure ejectors with the extracellular amplifier EXT-02F/1 in one 19″ instrument that can easily be mounted in a rack.

Last Updated on May 27, 2021

Details

Pressure ejection of substances can be performed with the same pipette as extracellular recording at the same position, or at a distant site using a separate pipette.

Specifications and options are the same for the ejection and recording part of the PEXT-01 as for the single channel instruments of the PDES and EXT-02F.

Features:

  • 2 or 4 bar pneumatic output
  • Digital manometer with PRESSURE MONITOR
  • Digital timer (100 ìs to 9999.9 ms) for pressure ejection
  • Manual, internal timed or external (TTL) pneumatic operation
  • Differential amplifier INPUT with high impedance
  • INPUT with BNC connectors or small headstage
  • OFFSET control potentiometer with two ranges (±0.1V or ±1V)
  • LOWPASS filter: 30, 100, 300, 500, 700, 900, 1.3k, 1.5k, 1.7k, 3k, 10k, 20k Hz
  • HIGHPASS filter: OFF, 3, 30, 100, 300, 400, 500, 600, 800, 1k, 3k, 10k Hz
  • OUTPUT: DC or AC FILTERED x10 to x10k
  • Optional AUDIO monitor, noise mode with threshold or pitch mode
  • OVER LEDs (activating at 10% below maximum output)
  • Analog BALANCE monitor for optimal OFFSET control
  • Two separate holders for ejection and recording or one combined ejection/recording pipette holder

Last Updated on May 27, 2021

Specifications - PDES

MODE SELECTION: 

three position toggle switch, (TIMER, CONT., EXT.).

SOURCE GAS: 

air, nitrogen, helium etc. no aggressive, combustible or explosive gases, filtered by a 5 µm filter element.

PRESSURE RANGE: 
input pressure: max. 5 bar (1 bar = 100 kPa = 14.5 psi)
pressure control range: 0-4 bar, 0-58 psi; with microJECT 0-2 bar, 0-30 psi
display range: 0-4 bar, 0-58 psi
residual air pressure: 10 mbar (0.145 psi) – 100 mbar (1.45 psi).

INPUT AND OUTPUT COUPLINGS: 
quick connect type.

DIGITAL TIMER: 
100 µs – 9.9 s set by five digit tumble-wheel switch
(0000.1 – 9999.9 ms).

ANALOG TIMER: 
0.1 – 1 s and 1 – 10 s with linear control, range set by toggle switch time base selector (0.1 s, 1 s).

MINIMUM PRESSURE PULSE LENGTH: 
10 ms (digital timer), <1 ms with microJECT, 100 ms (analog timer).

PULSE OUTPUT: 
Pressure Monitor: standard BNC connector, sensitivity: 1 V / bar or 0.1 V / psi, output impedance 249 Ω
Timing Monitor: standard BNC connector, TTL (LO = 0-1 V, HI = 5 V), output impedance 249 Ω.

PULSE INPUT: 
standard BNC connector, TTL (LO = 0-2.5 V, HI = 2.5-12 V), input impedance 10 kΩ.

STATUS / EJECT TIME INDICATOR: 
red/green LED
red light: channel has been turned on, valve is closed
green light: valve is open.

POWER REQUIREMENTS: 
115/230 V AC, 20-40 W,

DIMENSIONS: 
19″ rackmount cabinet, 19″ (483 mm), 10″ (250 mm), 3.5″ (88 mm)

REMOTE CONTROL (OPTION): 
Toggle switch and status LED for each channel,
size: 25 x 40 x 100 mm.

Last Updated on May 27, 2021

Specifications - EXT

HEADSTAGES (optional):

Operating voltage ±12 V; Size (approx.): 65x25x25mm
holding bar diameter 9 mm, length 150 mm
Electrode connector: BNC with driven shield
Ground connector: 2.4 mm connector or headstage enclosure
Reference connector: gold-plated SMB, grounded shield
Input resistance: >1012 Ω
Differential input: cmr >90 dB at 1 kHz (tested with 0 Ω input resistance).

AC/DC OUTPUT: 
filtered and amplified
impedance 50 Ω

DIFFERENTIAL INPUT: 
BNC / SMB connector with headstage or BNC connectors
CMR >90 dB
GND position for each INPUT
INPUT RESISTANCE (without headstage):
10E+6 Ω , range ±1 V (lower input resistance on request)

LOW PASS FILTER: 
30, 100, 500, 700, 900, 1.3k, 1.5k, 1.7k, 3k, 10k, 20k Hz

HIGH PASS FILTER: 

OFF, 3, 30, 100, 300, 400, 500, 600, 800, 1k, 3k, 10k Hz

NOTCH FILTER (optional): 
50 Hz or 60 Hz

AUDIO MONITOR: 
PITCH or NOISE, DC or AC
NOISE mode with THRESHOLD
channel selected by switch
connector external speaker / line

GAIN: 
10, 20, 50, 100, 200, 500, 1k, 2k, 5k, 10k

OFFSET COMPENSATION: 
Potentiometer with 2 ranges: ±0.1 V or ±1 V

MONITORS: 
Impedance: 250 Ohms
GAIN: BNC connector, 1 V per step, +1 V….+7 V
FILTER: BNC connector, 1 V per step, -8 V….+7 V
COMMAND: BNC connector, sensitivity: x10

ANALOG BALANCE MONITOR: 

ca. ±30 mV

EXTERNAL POWER SUPPLY: 
AC output; 12-18V, 1 A

DIMENSIONS: 

245 x 200 x 90 mm

Last Updated on May 27, 2021

Downloads

Manual:
PEXT-01 Manual

Brochure:
PEXT-01 Brochure

Last Updated on May 27, 2021

Ordering

Part No. Description
PEXT-01D PEXT System with Digital Timers, one Channel PDES w. Analog Manometer, EXT Amplifier w/o Headstage
PEXT-01D-LA PEXT System with Digital Timers, one Channel PDES w. Digital Manometer, EXT Amplifier w/o Headstage
PEXT-01DE PEXT System with Digital Timers, one Channel PDES w. Analog Manometer, requires external Valve, EXT Amplifier w/o Headstage
PEXT-01DE-LA PEXT System with Digital Timers, one Channel PDES w. Digital Manometer, requires external Valve, EXT Amplifier w/o Headstage

Last Updated on May 27, 2021

Quote List

No products in the list

Related products

  • QMM MicroManifold® Series

    Show Details
  • Pressure Ejection Systems

    Show Details
  • Programmable Single-Syringe Microfluidics Pump

    Show Details
  • Tubing

    Select options This product has multiple variants. The options may be chosen on the product page

Subscribe to our Newsletter


About Us

ALA Scientific Instruments, Inc. began as Adams and List Associates Ltd. in 1986.

Recent News

  • ALA is attending Society for NeuroScience
  • ALA is attending Biophysical Society Meeting
  • FENS FORUM 2024
  • ALA is attending Biophysical Society Meeting
  • ALA is attending Society for Neuroscience

Search Site

Copyright © 2025 - Ala Scientific Instruments, Inc.
  • Facebook
  • Twitter
  • Instagram
  • LinkedIn
  • Rss
Budget Friendly Dual Channel Extracellular Amplifier Voltammetry / Amperometry Amplifier
Scroll to top

This site uses cookies. By continuing to browse the site, you are agreeing to our use of cookies.

Accept settingsHide notification onlySettings

Cookie and Privacy Settings



How we use cookies

We may request cookies to be set on your device. We use cookies to let us know when you visit our websites, how you interact with us, to enrich your user experience, and to customize your relationship with our website.

Click on the different category headings to find out more. You can also change some of your preferences. Note that blocking some types of cookies may impact your experience on our websites and the services we are able to offer.

Essential Website Cookies

These cookies are strictly necessary to provide you with services available through our website and to use some of its features.

Because these cookies are strictly necessary to deliver the website, refuseing them will have impact how our site functions. You always can block or delete cookies by changing your browser settings and force blocking all cookies on this website. But this will always prompt you to accept/refuse cookies when revisiting our site.

We fully respect if you want to refuse cookies but to avoid asking you again and again kindly allow us to store a cookie for that. You are free to opt out any time or opt in for other cookies to get a better experience. If you refuse cookies we will remove all set cookies in our domain.

We provide you with a list of stored cookies on your computer in our domain so you can check what we stored. Due to security reasons we are not able to show or modify cookies from other domains. You can check these in your browser security settings.

Other external services

We also use different external services like Google Webfonts, Google Maps, and external Video providers. Since these providers may collect personal data like your IP address we allow you to block them here. Please be aware that this might heavily reduce the functionality and appearance of our site. Changes will take effect once you reload the page.

Google Webfont Settings:

Google Map Settings:

Google reCaptcha Settings:

Vimeo and Youtube video embeds:

Accept settingsHide notification only